FUGUANG

Hotline:400-668-8627丨MALL中文

1 1
Your location :
Homepage
/
/
/
HMA-TNI Total Nickel Heavy Metal Analyzer

Brand instruments

Brand Instrument

image/0kLMjQ2uTh-7jWZTCk5WNA

HMA-TNI Total Nickel Heavy Metal Analyzer

image/0kLMjQ2uTh-7jWZTCk5WNA
Product description
Technical indicators
-Measuring principle:
 
The instrument is detected by colorimetric spectrophotometry, the water sample is digested by high temperature, and the digestion reagent is added to the water sample. The sample is pumped to the high temperature digestion unit. The total nickel in the sample is digested into divalent nickel ions. Divalent nickel ions form an orange-brown colored complex with butadione oxime in an alkaline solution under the environment of an oxidant (ammonium persulfate). Spectrophotometric measurement was carried out at a wavelength of 470 nm. According to the initial color of the sample and the color after adding the developer, the difference between the two was analyzed to analyze the concentration of the sample.
 
 
 
- typical application:
 
On-line monitoring of total nickel in surface water, industrial wastewater (such as electroplating nickel-containing wastewater), and inlet and outlet of sewage plants
 
 
 
-Instrument features:
 
● Low waste liquid. The amount of waste liquid is only 1.9L / month
 
● Low reagent consumption. No need to change reagents in one month
 
● Automatic range, automatic dilution. Both low-range and high-range modes
 
● Automatic backwash sample filter, automatic cleaning before and after each test
 
● Automatic calibration
 
● 5.7 ”color LCD touch screen
 
● Chinese menu
 
Scan the QR code to read on your phone
We could not find any corresponding parameters, please add them to the properties table
Technical index
Measurement methods: Colorimetry
Measuring range: 0.05 ~ 3 (low), 0.4 ~ 10mg / L (high)
Detection limit: 0.01mg/l
Accuracy: ±3%F.S
Repeatability: ±3%F.S
Zero drift: ±5%F.S
measure time: 50min
Digestion time: 120 ℃ high temperature digestion
Automatic calibration cycle: Closed, 1 day, 2 days ... 7 days optional
Maximum reagent dosage: 5L / month
Sample flow rate: 1 ~ 3L / min, take 50mL each time for measurement
Sample temperature range: 10-40℃
Signal output: 4-20mA/RS232/RS485
Installation Environment: Indoor installation
power supply: AC220V±10%,50Hz
Dimensions (H × D × W): 900 × 255 × 600mm (analyzer body)
weight: 60kg (without reagents)

 

Customer message
 Fuguang

Mainly engaged in the distribution of environmental monitoring instruments and meters, as well as the integration and operation and maintenance of environmental monitoring systems; Research and development, production, and sales of environmental monitoring instruments;

Copyright ©2023 Fuzhou Fuguang Water Science&Technology    闽ICP备12013341号-2  Powered by saa 

 Fuguang

Follow Fuguang Water

 Fuguang

Official Taobao Store

Fuzhou Fuguang Water Science&Technology Co.,Ltd

闽ICP备12013341号-2  Powered by saa