Polymetron 9582 dissolved oxygen analyzer

Brand Instrument

Brand Instrument
 ${ele.title}
 ${ele.title}

Polymetron 9582 dissolved oxygen analyzer

Category: Polymetron
-Typical applications are power plants, semiconductor plants, etc., suitable for the analysis of dissolved oxygen at low concentrations (ppb) such as deaerated water. -Characteristics and advantages Pre-installed membrane modules, membrane replacement work only takes a few seconds, low detection limit, fast calibration process: less than 15 minutes to complete the calibration from ppm level (air) to ppb level (process), versatility 9500 controller, easy to integrate and simple to install. Working principle The measurement of dissolved oxygen is based on the well-known Clark battery principle. The oxygen-permeable membrane isolates the water sample from the electrode, so no sample adjustment is required. Due to other reducing or oxidizing ions.
Product description
Parameters
Technical indicators
- typical application
 
Power plants, semiconductor plants, etc., are suitable for the analysis of dissolved oxygen at low concentration (ppb) such as deaerated water.
 
-Features and benefits
 
The pre-installed membrane module only needs a few seconds to replace the membrane
 
Low detection limit
 
Quick calibration process: calibration from ppm (air) to ppb (process) can be completed in less than 15 minutes
 
The universal 9500 controller is used for easy integration and simple installation.
 
 
 
- working principle
 
The measurement of dissolved oxygen is based on the well-known Clark battery principle. The oxygen-permeable membrane isolates the water sample from the electrode, so there is no need to adjust the sample. Since other reducing or oxidizing ions cannot pass through the gas-permeable membrane, they will not cause interference.
 
A constant voltage is applied to the two electrodes to keep the potential of each electrode constant. The gold working electrode (cathode) reduces dissolved oxygen to hydroxide ions:
 
O2 + 2H2O + 4e--> 4OH
 
The oxidation reaction takes place on the surface of the large silver electrode (anode):
 
4Ag + + 4Br--> 4AgBr + 4e-
 
The oxidation-reduction reaction produces a diffusion current, which is proportional to the dissolved oxygen concentration.
 
We could not find any corresponding parameters, please add them to the properties table
Technical index
Measuring range 0 ~ 2000 ppb(0-2 ppm)
Minimum detection limit < 1 ppb
unit mg/L, ppm, μg/L, ppb, mbar, hPa,inch Hg, mmHg
Repeatability ± 0.5 ppb or ± 2%, whichever is greater
Reproducibility ± 0.5 ppb or ± 2%, whichever is greater
Response time For 1-40 ppb step change the value is: <30 seconds
Maintenance interval Membrane life: 6 months, depending on the sample
range of working temperature -20 ~ 60 ° C (0 ~ 95% RH, non-condensing)
Temperature compensation Automatic compensation within 0 ~ 45 ° C (32 ~ 113 ° F)
connection 1/4 inch NPT threaded connection (6mm or 1/4 inch pipe is recommended)
Sample flow grade 66 ~ 333 mL/min(4 ~ 10 L/hr)

 

FUGUANG

Mailing Address: Building B, No. 168 Jinyan Road, Jianxin Town, Cangshan District, Fuzhou

Postal Code: 350007

Customer Service Hotline: 400-668-8627

Main Line/Fax: 0591-83305533/83305866

Email: fgwater@fuguang.com

FUGUANG

Follow us

FUGUANG

Taobao store

Copyright © Fuzhou Fuguang Water Science & Technology Co.,Ltd. All rights reserved | Powered by 233 Technology