FUGUANG

Hotline:400-668-8627丨MALL中文

1 1
Your location:
Homepage
/
/
/
Surface Scatter 7sc High Range Turbidity Meter

Online Analyzer

Online Analyzer

Surface-Scatter-7sc-高量程浊度仪

Surface Scatter 7sc High Range Turbidity Meter

Typical applications are used to measure the turbidity of raw water, sedimentation tank effluent, and water with a high content of suspended matter. Suitable for drinking water / sewage / industrial water treatment industry. Instrument Features ● Optical components never touch water-no need to frequently clean the probe and other maintenance work ● Wide measurement range-can measure turbidity of 0 ~ 9999NTU ● Reliable and accurate measurement-the use of patented SurfaceScatter technology reduces interference ● Use Corrosion-resistant material can be used under high temperature sample conditions ● Equipped with reference calibration plate, which is convenient for understanding the current status of the instrument ● With standard controller sc200 or sc10
Surface-Scatter-7sc-高量程浊度仪
Product description
Technical indicators
typical application
It is used to measure the turbidity of raw water, sedimentation tank effluent and water with a high content of suspended matter. Suitable for drinking water / sewage / industrial water treatment industry.
 
Instrument characteristics
● The optical components never come into contact with water-no maintenance work such as frequent probe cleaning is required
● Wide measuring range-can measure turbidity of 0 ~ 9999NTU
● Reliable and accurate measurement-using patented Surface Scatter technology reduces interference
● Use anti-corrosion material, can be used under high temperature sample conditions
● Equipped with reference calibration board, which is convenient for understanding the current status of the instrument
● Connect with standard controller sc200 or sc1000
 
Measuring principle
Surface Scatter 7 sc is a turbidity meter that uses the principle of scattered light monitoring. The sample flows into the inclined turbidimeter body at a flow rate of 1 to 2 liters per minute, and the liquid overflows the horizontal surface at the top of the turbidimeter body to form a flat measurement plane. A strong light beam hits the liquid surface at an angle. The suspended particles in the sample emit scattered light. A detector placed above the liquid surface detects 90 degrees of scattered light.
Scan the QR code to read on your phone
We could not find any corresponding parameters, please add them to the properties table
Technical index
Measuring range 0~9999NTU
Accuracy At 0 ~ 2000NTU, ± 5% of reading
Sample flow rate 1.0~2.0L/min
Power requirements 220 VAC, 50 Hz, 40 VA
Sample temperature
High temperature 0-70 ℃, ordinary 0-50 ℃
Analog, digital and alarm signals: please refer to the technical specifications of sc200 or sc1000 standard controllers.

 

Customer message
 Fuguang

Mainly engaged in the distribution of environmental monitoring instruments and meters, as well as the integration and operation and maintenance of environmental monitoring systems; Research and development, production, and sales of environmental monitoring instruments;

Copyright ©2023 Fuzhou Fuguang Water Science&Technology    闽ICP备12013341号-2  Powered by saa 

 Fuguang

Follow Fuguang Water

 Fuguang

Official Taobao Store

Fuzhou Fuguang Water Science&Technology Co.,Ltd

闽ICP备12013341号-2  Powered by saa