FUGUANG

FUGUANG

/
/
/
EZ3000 Series Online Chloride Analyzer

Online Analyzer

Online Analyzer
 ${ele.title}
 ${ele.title}

EZ3000 Series Online Chloride Analyzer

Product description
Parameters
Technical indicators

Product Overview:

The EZ3000 series online chloride analyzer uses advanced sequential batch direct ion selective electrode (ISE) method, strictly adhering to EPA 9212 and ASTM D512-12 standard methods. Through an intelligent automatic control system and low-maintenance design, this instrument ensures data accuracy and reliability while significantly reducing operation and maintenance costs, providing efficient and continuous monitoring for water quality safety and compliant discharge.

 

Measurement Method:

The EZ3000 series online chloride analyzer adopts a sequencing batch direct ion selective electrode method, complying with standard methods EPA 9212 and ASTM D512-12, ensuring measurement accuracy and reliability.

 

Instrument Features:

1) The chloride detector uses a sequencing batch direct ISE analysis method, providing excellent analytical performance and ensuring the accuracy of detection results

2) This chloride detector requires minimal maintenance and consumes little reagent

3) The online chloride analyzer offers flexible measurement ranges and multiple selectable measurement ranges to meet different application needs

4) The online chloride analyzer is equipped with an intelligent automatic control system, simplifying the operation process and improving work efficiency

5) The chloride detector is controlled and communicated via an industrial panel computer, achieving convenient operation and visualized management

6) The online chloride analyzer offers both analog and digital output options, facilitating integration with other systems

7) The chloride detector supports multi-channel analysis (8 channels optional), suitable for high-throughput testing needs

 

Application Areas:

The EZ3000 series online chloride analyzers are widely used in drinking water and surface water, ensuring water quality safety and compliance with environmental protection standards.

We could not find any corresponding parameters, please add them to the properties table
Technical Specifications
measurement parameters chloride
measurement method sequential batch direct ion selective electrode method, complies with standard methods EPA 9212 and ASTM D512-12
standard measurement range EZ3003:1 - 10 mg/L Cl-
EZ3004:10 - 100 mg/L Cl-
EZ3005:100 - 1,000 mg/L Cl-
minimum detection limit ≤ 1 mg/L
accuracy ≤ 2% F.S. (using standard solution).
measure the interval 5 minutes
verification automatic, frequency can be set
calibration automatic calibration at two points, frequency can be freely set
operating temperature 5-35°C±5°C (avoid direct sunlight and rain), relative humidity 5-95% (non-condensing).
reagent requirements Store between 10-30℃
flow rate 100 - 300 mL/min
sample temperature 10℃ - 30℃
power requirements 110-240VAC, 4A, 50/60Hz, maximum power consumption: 150VA
simulated output 4-20mA, maximum 500Ohm load, standard 1 channel, up to 8 channels (optional).
digital output Optional: RS232, Modbus (TCP/IP, RS485).
sample requirements maximum particle size 100μm, <0.1 g/L; Turbidity < 50 NTU
verification complies with CE/UL certification
enclosure protection Analyzer host: IP55;
PC panel: IP65
size 690mmx 465mmx 330mm
shipping weight 30kg
FUGUANG

Mailing Address: Building B, No. 168 Jinyan Road, Jianxin Town, Cangshan District, Fuzhou

Postal Code: 350007

Customer Service Hotline: 400-668-8627

Main Line/Fax: 0591-83305533/83305866

Email: fgwater@fuguang.com

FUGUANG

Follow us

FUGUANG

Taobao store

Copyright © Fuzhou Fuguang Water Science & Technology Co.,Ltd. All rights reserved | Powered by 233 Technology