EZ7200 Volatile Fatty Acid Analyzer

Product description
Parameters
Technical indicators

Product Overview:

HACH EZ7200 is an on-line analyzer dedicated to anaerobic treatment scenarios. It directly measures VFA (volatile fatty acids), bicarbonate, total alkalinity and partial alkalinity by acid-base titration. It supports 8-channel multi-flow path analysis, can continuously monitor and prevent digestion tank failures caused by VFA accumulation, and can also integrate enterprise network convenient management and control, adapt to anaerobic wastewater treatment, sludge treatment and pilot reaction tank, and provide data support for stable operation of anaerobic process.

 

Working Principle:

  • Measurement methods

    The EZ7200 online volatile fatty acid (VFA) analyzer uses an acid-base titration method.

  • Measurement parameters

    The EZ7200 online volatile fatty acid (VFA) analyzer can measure volatile fatty acids (VFAs), bicarbonates, total alkalinity, and partial alkalinity.

 

Instrument Features:

1) The EZ7200 volatile fatty acid analyzer uses an acid-base titration method, allowing direct titration, resulting in low evaporation

2) Continuous monitoring of anaerobic processes to prevent digester failures caused by VFA accumulation

3) Easy-to-implement dynamic control strategies, combined with online monitoring, and the volatile fatty acid analyzer ensures the anaerobic process of the process

4) The EZ7200 online VFA monitor can be easily integrated into the company network, making operation simple

5) The EZ7200 online VFA monitor can achieve multi-stream analysis (8 channels)

 

Application Areas:

The EZ7200 online volatile fatty acid (VFA) monitor can be applied to anaerobic wastewater treatment, anaerobic sludge treatment, and pilot-scale anaerobic reaction tanks.

We could not find any corresponding parameters, please add them to the properties table
Technical Specifications
Measurement parameters Volatile fatty acids (VFAs), bicarbonates, total basicity, partial basicity
Measurement method Acid-base titration method
Measurement range VFA:10-10,000 mg/L Calculated as acetic acid
Bicarbonate: 0-100 meq/L or 10,000 mg/L CaCO3
Total and partial basicity: 0-100 meq/L or 10,000 mg/LCaCO3
Check the limit value ≤ 10 mg/L (range 10 - 500 mg/L VFA).
Accuracy ≤3% F.S. (using standard solution).
Measure the interval 10-15 minutes
Verification Automatically verifies, frequency can be set
Calibration Auto-calibration, frequency can be set
Ambient temperature 10°C - 30°C±4°C, relative humidity: 5%-95% (non-condensing).
Reagent requirements Store between 10 and 30 ℃
Sample flow rate 100-300ml/min
Sample temperature 10℃ - 30℃
Instrument air According to ISA-S7.0.01-1996 Instrumented Air Quality Standard, oil-free drying is required
Power requirements 110 - 240 VAC, 4 A, 50/60 Hz, maximum power consumption: 150VA
Simulated output 4-20 mA, maximum 500-ohm load, standard 1 analog output, up to 8 channels (optional).
Digital output Optional: RS232, Modbus (TCP/IP, RS485).
Sample requirements Maximum particle size 500μm, <0.1 g/L            
Verification Complies with CE/UL certification
Enclosure protection Analyzer host: IP55;
PC panel: IP65
Size 690mmx 465mmx 330mm
Shipping weight 25kg
FUGUANG

Mailing Address: Building B, No. 168 Jinyan Road, Jianxin Town, Cangshan District, Fuzhou

Postal Code: 350007

Customer Service Hotline: 400-668-8627

Main Line/Fax: 0591-83305533/83305866

Email: fgwater@fuguang.com

FUGUANG

Follow us

FUGUANG

Taobao store

Copyright © Fuzhou Fuguang Water Science & Technology Co.,Ltd. All rights reserved | Powered by 233 Technology