Product description
Parameters
Technical indicators

Working Principle:

The Surface Scatter 7sc high-range turbidity meter uses the principle of scattered light monitoring. The sample flows into the tilted turbidity meter body at a rate of about 1 to 2 liters per minute, with the liquid overflowing onto the horizontal surface at the top of the turbidity meter body, forming a flat measurement plane. A strong beam of light is directed at the liquid surface at a certain angle, causing suspended particles in the sample to emit scattered light. A detector placed above the liquid detects the 90-degree scattered light, completing turbidity measurement.

 

Application Industries:

The Surface Scatter 7sc high-range online turbidity meter is widely used in the drinking water, wastewater, and industrial treated water industries, especially suitable for measuring turbidity in water samples with high suspended solids content, such as raw water and sedimentation tank effluent.

 

Instrument Features:

● This high-range turbidity meter uses Surface Scatter technology to reduce interference and ensure reliable and accurate turbidity measurement.

● The Surface Scatter 7sc high-range turbidity meter features optical components that do not come into contact with water, eliminating the need for frequent probe cleaning and reducing maintenance.

● High-range online turbidity meter has a wide range, with a wide range of 0~9999 NTU, suitable for various complex water quality environments.

● The Surface Scatter 7sc high-range turbidity meter uses corrosion-resistant materials and can detect high-temperature samples.

● Surface Scatter 7sc high-range turbidity meter comes with a reference calibration plate, making it easier to understand the current status of the instrument.

● The Surface Scatter 7sc high-range turbidity meter can connect with standard controllers sc200 or sc1000 for smarter control and data management.

We could not find any corresponding parameters, please add them to the properties table
Technical Specifications
Measurement range: 0~9999NTU
Accuracy: At 0~2000 NTU, the reading is ±5%.
Or ±0.1 NTU, whichever is larger;
At 2000~9999 NTU, the reading ± is 10%.
Sample flow rate: 1.0~2.0L/min
Power requirements: 220 VAC, 50 Hz, 40 VA
Sample temperature: High temperature  0-70°C, normal 0-50°C
Analog, digital, alarm signals: Please refer to the technical specifications of the SC200 or SC1000 standard controllers
FUGUANG

Mailing Address: Building B, No. 168 Jinyan Road, Jianxin Town, Cangshan District, Fuzhou

Postal Code: 350007

Customer Service Hotline: 400-668-8627

Main Line/Fax: 0591-83305533/83305866

Email: fgwater@fuguang.com