The QL3580I Condensate Total Organic Carbon (TOC) analyzer is a precision instrument specifically designed for high-purity water quality monitoring in industries such as power, chemical, and metallurgy. This product uses UV persulfate oxidation and NDIR detection technology to accurately determine the TOC content in condensate water, effectively monitor water quality in thermal systems, prevent corrosion and scaling caused by organic pollutants, and ensure long-term stable operation.

 

Measurement Method:

QL3580I condensate total organic carbon detector uses advanced measurement methods, with the following specific steps:

1. The water sample enters the reactor and reacts with acid, converting the inorganic carbon in the sample into carbon dioxide;

2. Using air without carbon dioxide (instrument air) as the carrier gas to discharge carbon dioxide converted into inorganic carbon; at this point, the NDIR detector can detect total inorganic carbon (TIC);

3. Then oxidize the water sample with TIC (total inorganic carbon) removed;

4. After oxidation is completed, the carbon dioxide converted into organic carbon is discharged through carrier gas and detected by an NDIR detector to determine the total organic carbon (TOC) content.

 

Instrument Features:

  • High-precision oxidation and detection technology

    Using UV persulfate oxidation to ensure complete conversion and accurate measurement of organic carbon, with stable and reliable data.

  • Anti-clogging injection design

    Equipped with a 4mm inner diameter sample inlet pipeline, capable of tolerating 2mm soft particles; conventional water samples require no pretreatment, reducing maintenance frequency.

  • Fully automatic calibration function

    Supports two-point automatic calibration, reducing manual intervention and lowering operational errors and maintenance workload.

  • User-friendly human-machine interface

    Equipped with a high-definition touchscreen, it supports multi-level permission management, facilitating data export, software updates, and daily operations.

  • Low reagent consumption operating mode

    Sequential batch measurement optimizes reagent usage, combined with alkaline-free reagent design to reduce long-term operating costs.

  • Compact and flexible structural design

    Compact in size, supporting multiple installation methods such as wall mounting and brackets, suitable for explosion-proof cabinet applications, space-saving, and suitable for harsh environments.

 

Application Areas:

QL3580I condensate total organic carbon analyzers are generally used in industrial condensate testing projects to ensure water quality monitoring meets industry standards. It is widely used in the quality control and management of condensate in fields such as power, chemical, and metallurgy, ensuring water safety and long-term stable equipment operation.

We could not find any corresponding parameters, please add them to the properties table
Technical Specifications
Measurement parameters TOC (NPOC)/TIC
Oxidation method UV persulfate oxidation
Measurement method After oxidation, carbon dioxide concentration is detected using an NDIR detector
Sample submission requirements Flow rate: ≥100 mL/min
Particle size: ≤2 mm, soft particle
Injection pressure: ≤1 bar
Carrier gas Instrument wind with CO2 removal:
Source pressure: 50~120 PSIG
Dew point: -20 °C (removes water vapor and oil vapor)
Instrument gas consumption flow rate: maximum 1.7 L/min
Water sample temperature 5 ℃ - 45 ℃
Analysis time 7 minutes (TOC).
Measure the interval Adjustable (in minutes).
Repeatability TOC:±0.5 mg C/L or ±2% of the reading (whichever is greater); TIC: ±4% of the reading
Accuracy ±0.5 mg C/L or ±3% of the reading, whichever is greater
Interference Possible turbidity, fat, oil, protein, tar can interfere with measurement results
Protection level IP44 (with case).
Signal output 0/4~20 mA
Relay output 2 relays (24V 4A dry contact).
Display 7-inch, 800 × 480 pixels (4:3) TFT-LCD touchscreen
Data storage 9,999 analysis data points, 99 fault information
Power supply 220V AC, 50 Hz
Size 750 × 500 × 320 mm
Weight 55 Kg
FUGUANG

Mailing Address: Building B, No. 168 Jinyan Road, Jianxin Town, Cangshan District, Fuzhou

Postal Code: 350007

Customer Service Hotline: 400-668-8627

Main Line/Fax: 0591-83305533/83305866

Email: fgwater@fuguang.com

FUGUANG

Follow us

FUGUANG

Taobao store

Copyright © Fuzhou Fuguang Water Science & Technology Co.,Ltd. All rights reserved | Powered by 233 Technology