9184SC residual chlorine online detector

Product description
Parameters
Technical indicators

Product Overview:

The HACH 9184sc residual chlorine online detector uses selective electrodes for residual chlorine detection, featuring plug-and-play functionality, a wide range, and low detection limits; This residual chlorine online detector has built-in pH/temperature automatic compensation, making it especially suitable for residual chlorine detection in membrane treatment processes such as reverse osmosis, ensuring more accurate readings.

 

Working Principle:

The principle of the 9184SC residual chlorine online detector is: the electrolyte and the permeation membrane separate the electrolytic cell from the water sample, allowing the permeation membrane to selectively allow ClO- penetration; There is a fixed potential difference between the two electrodes, and the generated current intensity can be converted into residual chlorine concentration. Through the following reactions, the system can accurately measure residual chlorine:

At the cathode: ClO- + 2H+ + 2e- →Cl- + H2O

On the anode: Cl- + Ag→ AgCl + e-

Due to the fixed conversion relationship between HOCl, ClO-, and residual chlorine under specific temperature and pH conditions, the 9184SC residual chlorine detector can accurately detect residual chlorine using the electrode method.

 

Application Industries:

The 9184SC residual chlorine online detector is suitable for detecting hypochlorous acid (HOCL)/residual chlorine in drinking water and industrial water disinfection processes, especially for detecting residual chlorine in influent water from membrane treatment processes such as reverse osmosis.

 

Instrument Features:

● 9184sc residual chlorine online detector has wide range and low detection limit, meeting the needs of low range measurement users.

● 9184 sc residual chlorine online detector uses selective membrane sensor to ensure the accuracy of measurement.

● 9184sc residual chlorine online detector built-in pH and temperature compensation, so that the reading is closer to the true value.

● 9184sc portable chlorine detector built-in flow control device to ensure stable measurement.

● Low maintenance: calibrated every two months, and electrolyte and membrane replaced every six months.

● Plug and play: Uses the electrode method; when using, just insert the electrode and power on to perform residual chlorine detection.

We could not find any corresponding parameters, please add them to the properties table
Technical Specifications
Measurement range: 0.005 - 20 ppm (mg/L) HOCl
Minimum detection limit: 5 ppb or 0.005 mg/L HOCl
Accuracy: 2% or ± 10 ppb  HOCl, whichever is higher
Response time: 90% less than about 90 seconds
Sample flow rate: 200~250 mL/min Automatic adjustment
Storage temperature: -20~60℃
Operating temperature: 0~45℃
Sample temperature: 2~45℃
Sample pH range: 4 ~ 8
Calibration method: Laboratory comparison method
Calibration interval: One time/2 months
Maintenance interval: Generally, replace the membrane and electrolyte every six months
Feed Link: 1/4-in. O.D.
Row Link: 1/2-in. I.D.
Protection level: IP-66/NEMA 4X
Instrument size: 270 x 250 mm
FUGUANG

Mailing Address: Building B, No. 168 Jinyan Road, Jianxin Town, Cangshan District, Fuzhou

Postal Code: 350007

Customer Service Hotline: 400-668-8627

Main Line/Fax: 0591-83305533/83305866

Email: fgwater@fuguang.com

FUGUANG

Follow us

FUGUANG

Taobao store

Copyright © Fuzhou Fuguang Water Science & Technology Co.,Ltd. All rights reserved | Powered by 233 Technology