FUGUANG

FUGUANG

/
/
/
ULR CL17sc online total chlorine analyzer

Online Analyzer

Online Analyzer
 ${ele.title}
 ${ele.title}

ULR CL17sc online total chlorine analyzer

Product description
Parameters
Technical indicators

Measurement Method:

The total chlorine analyzer uses methods such as N,N-diethyl-1,4-phenylenediamine photometry (DPD) for total chlorine determination

 

Application Industries:

The ULR CL17sc online total chlorine analyzer is widely used in industries such as wastewater, food and beverage production, pharmaceuticals, power, and semiconductors.

 

Instrument Features:

The ULR CL17sc online total chlorine analyzer can accurately detect chlorine content above 8ppb in water, allowing monitoring and optimizing the dechlorination process.

The online total chlorine analyzer uses the DPD colorimetric method to directly measure the total chlorine content in water, making it convenient and fast.

The total chlorine analyzer features a cumulative chlorine content function, allowing customers to understand the true impact of chlorine content through reverse osmosis membranes, thereby more accurately predicting membrane efficiency and service life.

We could not find any corresponding parameters, please add them to the properties table
Technical Specifications
Measurement parameters Total chlorine
Measurement method Total chlorine: N,N-Diethyl-1,4-phenylenediamine photometry (DPD method).
Measurement range 0.008 - 5 mg/L Cl2
Accuracy ±5% or ±0.01mg/L (whichever is greater), 0-4mg/L;
±10%, 4-5mg/L
Check the limit 0.008 mg/L Cl2 , 25℃
Measure the period 2.5min
Inlet pressure 0.3 - 5.2 bar (4.5 - 75 psig), supplied to Y-type filter
0.1 - 0.3 bar (1.5 - 5 psig), supplied to the analyzer
Water inlet 1/4 inch outer diameter polyethylene pipe, quick connector
Drainage outlet 1/2 inch outer diameter hose
Air purging 3/8 inch quick connector and pipe fitting can be selected; At a maximum value of 1.38 bar, 0.003 m3/min
Sample requirements Y-type filters with mesh sizes of 40 mesh or higher
Flow rate The flow rate through the instrument is 60 - 200 mL/min
Sample temperature 5 - 40 °C
Operating temperature range 5 - 40 °C
Operating humidity 0-90% relative humidity, no condensation
Interference factors Oxidants such as bromide, chlorine dioxide, permanganate, and ozone can cause positive errors
Reagent consumption Use 0.5 L buffer solution and 0.5 L indicator solution within 31 days
Authentication CE compliant: EN 61326-1, CISPR 11, EN 50581
ACMA RCM
Korea KC certification
North America: FCC Supplier Declaration of Conformity, IEC/EN 60529, ICES-003
Light source LED, measured at 510 nm, optical path length 1 cm
Waterproof rating IP66
Installation Wall-mounted
Power supply requirements (voltage). 12 VDC, maximum 400 mA (powered by controller).
Size (height x width x depth). 342 mm x 329 mm x 177 mm
Weight 4.1 kg
FUGUANG

Mailing Address: Building B, No. 168 Jinyan Road, Jianxin Town, Cangshan District, Fuzhou

Postal Code: 350007

Customer Service Hotline: 400-668-8627

Main Line/Fax: 0591-83305533/83305866

Email: fgwater@fuguang.com

FUGUANG

Follow us

FUGUANG

Taobao store

Copyright © Fuzhou Fuguang Water Science & Technology Co.,Ltd. All rights reserved | Powered by 233 Technology