FUGUANG

FUGUANG

/
/
/
MS6100 Multi-Parameter Online Water Quality Analyzer

Online Analyzer

Online Analyzer
 ${ele.title}
 ${ele.title}

MS6100 Multi-Parameter Online Water Quality Analyzer

Product description
Parameters
Technical indicators

Product Overview:

As a multi-parameter online water quality analyzer, the MS6100 automatic water quality analyzer can continuously monitor seven parameters: residual chlorine, total chlorine, turbidity, pH, ORP, conductivity, and temperature. The turbidity module of this multi-parameter online water quality analyzer uses 360° x 90° detection technology, featuring precise measurement, low maintenance requirements, and flexible configuration. It is especially suitable for applications requiring efficient automated monitoring such as drinking water networks, secondary water supply, and pool water.

 

Working Principle:

The MS6100 integrates multiple advanced detection technologies and achieves precise multi-parameter monitoring through an intelligent sensing system:

  • Residual chlorine/total chlorine detection

    Using DPD photometry, chlorine in the water sample reacts with N,N-diethyl-1,4-phenylenediamine to form a red compound, which is determined by colorimetry.

  • Turbidity detection

    Using innovative 360° ×90° scattered light detection technology, scattered light signals are collected from all angles around the incident light beam, significantly improving measurement accuracy.

  • Electrochemical parameter detection

    pH/ORP/conductivity is electrochemically used, temperature is measured using NTC30K sensors, all with automatic temperature compensation.

 

Instrument Features:

  • Integrated design

    Integrated design with a small footprint. Continuous monitoring of seven water quality parameters, including residual chlorine, total chlorine, turbidity, pH, ORP, conductivity, and temperature.

  • Flexible configurations available

    Optional single/dual-channel chlorine modules to simultaneously monitor total chlorine and residual chlorine parameters. Optional electrochemical modules; pH/ORP/conductivity electrodes can be configured according to requirements.

  • Lower maintenance requirements

    Specifically designed for unattended applications. Water and power outages can be automatically protected and restored. If water or liquid leakage is detected internally, it can automatically cut off the water flow to prevent the instrument from being flooded. Chlorine measurement intervals are adjustable, ensuring reagent replacement cycles are at least once every 3 months, resulting in low reagent consumption.

  • Comprehensive communication functions

    Equipped with 2 RS485 interfaces (supporting MODBUS protocol and Hach SC protocol). Uses the standard RTU Modbus communication protocol. It also offers an optional 7-channel isolation output of 4-20 mA, making it convenient for users to transmit data to the data processing and monitoring center.

 

Application Areas:

The MS6100 multi-parameter online water quality analyzer can be applied in drinking water pipelines, secondary water supply, pool water, and other fields, providing continuous and accurate water quality monitoring for these scenarios.

  • Drinking water safety monitoring

    Continuous monitoring of water quality in municipal water supply networks and secondary water supply systems ensures drinking water safety.

  • Recreational water monitoring

    Real-time monitoring and automatic control of water quality for swimming pools, spas, and other recreational water.

We could not find any corresponding parameters, please add them to the properties table
Technical Specifications
Measurement parameters Residual chlorine, total chlorine, turbidity, pH, ORP, conductivity, and temperature
Measurement method Residual chlorine/Total chlorine: N,N-diethyl-1,4-phenylenediamine photometry (DPD method)
Turbidity: 90-degree scattering turbidimetric method, collecting scattered light around the incident beam at 360 degrees;
pH/ORP/ conductivity: electrochemical method;
Temperature: NTC 30K
Measurement range Residual chlorine/total chlorine: 0 ~ 5 mg/L as CL2;
Turbidity: 0 ~ 20 NTU;
pH:4.00 ~ 10.00;
ORP:-1200 mV ~ +1200 mV;
Conductivity: 20.0 μS/cm ~ 2000 μS/cm;
Temperature: 0 °C ~ 50 °C
Check the limit Residual chlorine/total chlorine: 0.03 mg/L;
Turbidity: 0.0032 NTU;
Conductivity: 6.0 μS/cm
Accuracy Residual chlorine/Total chlorine: ±5% or ±0.03 mg/L as CL2 (whichever is greater);
Turbidity: ±2% or ±0.01 NTU (whichever is higher); (Based on Formazin Grade 1 standard solution at 25°C)
pH:±0.02;
ORP:±20 mV;
Conductivity: ± 0.8% F.S.;
Temperature: ±0.4 °C
Repetition Residual chlorine/Total chlorine: 5% or 0.01 mg/L as CL2 (whichever is greater);
Turbidity: 1% or 0.006 NTU (whichever is greater); (Based on Formazin Grade 1 standard solution at 25 °C)
pH:0.01;
ORP:10 mV;
Conductivity: 0.4% F.S.;
Temperature: 0.2 °C
Response time Electrochemical electrode (pH electrode/ORP electrode/conductivity electrode) T90 response time< 3 minutes;
The entire T90 system response time (calculated from the moment the water sample enters the instrument) < 10 minutes.
Measure the interval Residual chlorine/total chlorine measurement interval configurable (continuous mode, 5 minutes, 10 minutes, 15 minutes, 20 minutes, 30 minutes);
Continuous measurement of turbidity, pH, ORP, conductivity, and temperature
Calibration method Residual chlorine / Total chlorine calibration: Factory calibrated, using default calibration curves;
Turbidity calibration: StabCal® single-point calibration (20 NTU), Formazin two-point calibration, custom calibration from 2 to 6 points;
pH calibration: single-point calibration, two-point calibration, three-point calibration;
Conductivity calibration: single-point calibration
Operating temperature 0 °C to +45 °C
Sample temperature 4 °C ~ 40 °C
Storage temperature Instrument: -20 °C to +55 °C;
Electrodes: -15 °C to +55 °C
Relative humidity 5% ~ 95% (no condensation).
Power requirements (220±20) VAC/50 Hz (factory default); Or (115±5) VAC/60 Hz (manual switching required);
Analog output 7 channel isolation 4-20 mA output
Output 2 RS485 interfaces (one of which supports MODBUS).
Protocol, another supporting Hach SC protocol).
Sample flow The instrument's sample injection flow path is split into two, and the user can adjust the sample flow rate for each flow path separately:
- Chlorine module flow regulation range: 120-180 ml/min (recommended: 165 ml/min);
- Turbidity (electrochemical) module flow regulation range: 150-450 ml/min (recommended: 250 ml/min).
Authentication CE certified;
ETL certification.
Enclosure protection IP65 (door with sealing strip, with door latch).
Size 49.0 cm (width) x 80.0 cm (depth) x 36.5 cm (thickness).
Shipping weight 45 kg (including packaging).
FUGUANG

Mailing Address: Building B, No. 168 Jinyan Road, Jianxin Town, Cangshan District, Fuzhou

Postal Code: 350007

Customer Service Hotline: 400-668-8627

Main Line/Fax: 0591-83305533/83305866

Email: fgwater@fuguang.com