Product description
Parameters
Technical indicators

Product Overview:

The MS5050 is a highly integrated panel-type online water quality analyzer, specifically designed for secondary water supply, rural drinking water, and similar scenarios. It integrates monitoring of multiple parameters such as residual chlorine/chlorine dioxide, pH, temperature, and turbidity, making installation simple and requiring low maintenance.

 

Working Principle:

  • 1. Residual chlorine/chlorine dioxide probe

    It uses a constant voltage method and is equipped with a three-electrode structure (working electrode, counter electrode, and reference electrode). Integrates pH glass electrodes and PT100 temperature electrodes, and uses built-in algorithms to dynamically compensate for interference with pH fluctuations and temperature changes in residual chlorine measurement in real time.

  • 2. Turbidity probe

    Based on the 90° scattering principle, the light source uses warm white LEDs (color temperature 3200K). Complies with USEPA 180.1 standards.

  • 3. ORP probe

    Based on a composite electrode structure of platinum electrodes and reference electrodes, it is used for accurate measurement of redox capacity in water.

  • 4. Conductivity

    Uses graphite four-electrode measurement technology, with built-in PT100 temperature compensation. It enables accurate conductivity measurement over a wide range (0-100 mS/cm).

 

Instrument Features:

● The online multi-parameter water quality detector uses composite residual chlorine/chlorine dioxide electrodes, enabling a single probe to output residual chlorine (chlorine dioxide), pH, and temperature parameters.

● Chlorine electrodes are made of gold material for greater stability.

● The electrode has built-in dual compensation functions for temperature and pH, using a unique algorithm to reduce the impact of lower temperature and flow rate on residual chlorine and pH value. This electrode achieves a film-free system without filler liquid; Used under drinking water conditions, requiring low maintenance.

● Turbidity sensor LED detection light source compliant with USEPA 180.1 standard color temperature and approved under USEPA 40 CFR 141.74(a)(1). The maximum resolution can reach 0.001 NTU, allowing long-term stable operation.

● The instrument is factory-calibrated, easy to install, simple to operate, and can be optionally equipped with conductive and ORP probes as needed.

● The instrument takes up little installation space, is easy to maintain, and requires minimal professional expertise from operators.

 

Application Areas:

The MS5050 panel-type multi-parameter online analyzer features an integrated design to meet customers' secondary water supply and rural drinking water monitoring needs.

We could not find any corresponding parameters, please add them to the properties table
Technical Indicators
standard probe parameters
project residual chlorine/chlorine dioxide PH temperature turbidity
measurement range 0-5 mg/L Cl2 or CLO2 count 2-12 -5℃~65℃ 0-10/10-40NTU
check the limit value 0.01 mg/L 0.01 pH -5℃ 0.001 NTU
display resolution 0.01 mg/L 0.01 pH 0.01℃ 0.001 NTU
accuracy ± 0.01 ppm or ±2% reading, whichever is greater ±0.02pH@25℃ ±1% ±0.01NTU or ±2% reading, whichever is higher
repeatability 2% 0.02 pH 0.5℃ 1%
minimum resolution 0.01 mg/L 0.01 pH 0.1℃ 0.001NTU
response time T90≤120S T90≤30S T90≤30S T90≤30S
measure the interval continuous measurement continuous measurement continuous measurement continuous measurement
joint protection level IP65 IP65 IP65 IP65
signal output 4-20mA and RS-485 dual outputs     4-20mA and RS-485 dual outputs
probe material SS304 housing, electrode tip made of gold + platinum + glass + CPVC + SS304 SS304 case + optical glass
Optional Probe Parameters
project ORP (optional). conductivity (optional).
measurement range ±1500mV 0-100,000μS/cm
check the limit value -1500 mV 10 μS/cm
display resolution 1 mV 1 μS/cm
accuracy ±20mV or 2% FS, whichever is larger ±10μS/cm or ±1% of the reading, whichever is higher
repeatability 2% 1%
minimum resolution 1 mV 1 μS/cm
response time T90≤25S T90≤20S
measure the interval continuous measurement continuous measurement
joint protection level IP65 IP65
signal output 4-20mA and RS-485 dual outputs 4-20mA and RS-485 dual outputs
probe material Housing: CPVC; Electrodes: Platinum Housing: CPVC; Electrode: graphite

 

System Parameters
project parameters
display 4.3" TFT 480x272 touchscreen
power supply 100-240VAC,50/60 Hz
power about 30W
analog output 4-20mA output each for residual chlorine, pH, turbidity, and temperature
communication interface RS485、Modbus RTU
data export USB drive
system upgrade USB drive
operating temperature 5-45℃
Storage temperature -20~55°C (excluding probe).
relative humidity 5% - 95% (no condensation).
altitude ≤2000 meters
water sample requirements temperature: 5~45°C
Pressure: 0.05~0.25 MPa, above 0.25 MPa, a pressure reducing valve
is required Flow rate of the circulation tank: greater than 470mL/min
cabinet material fully plastic, corrosion-resistant
size 750 mm×450 mm×230.5 mm
weight 15.2 KG (excluding packaging).
FUGUANG

Mailing Address: Building B, No. 168 Jinyan Road, Jianxin Town, Cangshan District, Fuzhou

Postal Code: 350007

Customer Service Hotline: 400-668-8627

Main Line/Fax: 0591-83305533/83305866

Email: fgwater@fuguang.com

FUGUANG

Follow us

FUGUANG

Taobao store

Copyright © Fuzhou Fuguang Water Science & Technology Co.,Ltd. All rights reserved | Powered by 233 Technology