Product description
Parameters
Technical indicators

Product Introduction:

The trace dissolved oxygen analyzer is an indispensable instrument in water quality analysis systems in the power industry. The HACH2582sc trace dissolved oxygen analyzer consists of an SC4500 controller and a market-proven trace dissolved oxygen sensor. This electrochemical trace dissolved oxygen analyzer uses a consistent product design platform, combined with familiar touchscreen design, skilled sensor assembly experience, and the same installation dimensions as the 9500 controller. The 2582SC is easier to install, integrate, and debug, and is especially suitable for precise measurement of low concentrations (ppb) dissolved oxygen.

 

Working Principle:

The dissolved oxygen measurement of the micro dissolved oxygen meter is based on the well-known Clark battery principle. The oxygen-permeable membrane isolates the water sample from the electrode, so there is no need to adjust the water sample.
Because their reducing or oxidizing ions cannot pass through the breathable membrane, they do not cause interference.
A constant voltage is applied to both electrodes, keeping the potential of each electrode constant.
The gold working electrode (cathode) reduces dissolved oxygen to hydroxide ions:
O2 + 2H2O + 4e- --> 4OH
Oxidation reactions occur on the surface of large silver electrodes (anodes):
4Ag+ + 4Br- --> 4AgBr + 4e-
Redox reactions produce diffusion current, and the magnitude of the current is proportional to the dissolved oxygen concentration. Electrochemical reactions and diffusion rates are sensitive to temperature. Therefore, the measurement unit is equipped with a built-in temperature sensor that can automatically compensate for temperature, improving measurement accuracy.

 

Product advantages of the 2582SC trace dissolved oxygen analyzer:

  • Simple integration and easy operation

    The trace dissolved oxygen analyzer is an indispensable instrument in the complete water quality analysis system of the power industry. Hash offers a wide range of water quality analyzer products, forming flexible integrated solutions that meet your unique needs and effectively save time on design, installation, training, maintenance, and operation.

  • Easy to master, saving time and effort

    The 2582sc trace dissolved oxygen analyzer consists of an SC4500 controller and a market-proven Hach trace dissolved oxygen sensor. A consistent product design platform that reduces customers' time searching for product information. The familiar touchscreen design, skilled sensor assembly experience, and the same installation dimensions as the 9500 controller ensure seamless integration and debugging of the 2582SC.

  • Simplified maintenance and operational processes

    The SC4500's universal menu guide is easy to operate and offers step-by-step maintenance and calibration procedures. Standard visual alarm signals can promptly notify operators for troubleshooting. Pre-installed film caps and factory-installed sensors minimize instrument startup and maintenance time.

  • Multiple communication methods are available

    The SC4500 controller provides local communication with SCADA or PLC. In addition to mA analog output, advanced digital communication protocols or LAN communication are available, allowing you to flexibly adapt to a rapidly changing world.

 

Application Areas:

This electrochemical trace dissolved oxygen analyzer is suitable for power plants, semiconductor factories, and similar locations for low-concentration (ppb) dissolved oxygen analysis such as deaerated water. This trace dissolved oxygen analyzer is suitable for this instrument.

 

Instrument Features:

● The electrochemical trace dissolved oxygen analyzer uses electrochemical analysis methods, which suits the habits of more users, making it classic and reliable.

● The micro dissolved oxygen detector has fast calibration speed, completing calibration from ppm to ppb level in less than 15 minutes.

● The trace dissolved oxygen analyzer has a low detection limit, with a minimum detection limit of < 1 ppb, meeting the needs of trace dissolved oxygen detection.

● The trace dissolved oxygen analyzer uses a universal SC4500 controller, facilitating integration with other systems.

● Dissolved oxygen probe installation is simple; membrane components are pre-installed and can be replaced in just a few seconds.

We could not find any corresponding parameters, please add them to the properties table
Technical Indicator
measurement range 0 - 2000 ppb (0 - 2 ppm)
Minimum Detection Limit (LOD) < 1 ppb
unit mg/L, ppm, μg/L, ppb, mbar, hPa, inch Hg, mmHg
repetition ±0.5 ppb or ±5%, take the larger value
reproducibility ±0.5 ppb or ±2%, take the larger value
response time 1-40 ppb:<30 seconds
calibration method zero: electron zero or oxygen-free water
Slope: Air calibration or calibration by comparison with laboratory instruments
maintenance cycle Dissolved oxygen membrane lifespan: ~6 months, depending on the sample
drain pipe 8mm pipeline
temperature compensation automatic, compensation range: 0 - 45°C
concatenate 1/4-inch NPT threaded connection (6mm or 1/4-inch piping recommended).
cable length 10m
monitor 3.5-inch TFT color monitor with capacitive touchscreen
protection level UL50E 4X model, IEC/EN 60529-IP 66, NEMA 250 4X, metal casing with corrosion-resistant
sample flow rate 66 - 166mL/min
operating temperature -20 ~ 60°C (8W(AC)/9W(DC) sensor load)
-20 ~ 45°C (sensor load of 28W (AC) / 20W (DC) )
Linear derating at 45 - 60°C (-1.33 W/°C)
storage conditions -20 - 70°C
power supply 100 - 240VAC±10%; 50/60Hz; 1A (28W sensor load)
24VDC +15% -20%; 2.5A (20W sensor load).
communication 5 x 0 - 20 mA or 4 - 20 mA analog output
Profibus DP, Modbus TCP, Profinet IO, Ethernet IP (optional upon request).
relay two relays (SPDT);
Gauge: 0.75 - 1.5 mm² (18 - 16 AWG)
AC controller
Maximum switching voltage: 100 - 240 vac
Maximum switching current: 5A resistor / 1A pilot
Maximum switching power: 1200VA resistor / 360VA pilot
DC controller
Maximum switching voltage: 30VAC or 42VDC
Maximum switching current: 4A resistor / 1A pilot
Maximum switching power: 125W resistor / 28W pilot frequency
Authentication CE, ETL certification compliant with UL and CSA safety standards (all sensor types), FCC, ISED, KC, RCM, EAC, UKCA,
SABS, CMIM, Morocco
Weight 3.2 kg
FUGUANG

Mailing Address: Building B, No. 168 Jinyan Road, Jianxin Town, Cangshan District, Fuzhou

Postal Code: 350007

Customer Service Hotline: 400-668-8627

Main Line/Fax: 0591-83305533/83305866

Email: fgwater@fuguang.com

FUGUANG

Follow us

FUGUANG

Taobao store

Copyright © Fuzhou Fuguang Water Science & Technology Co.,Ltd. All rights reserved | Powered by 233 Technology