-Typical applications are suitable for use in harsh environmental conditions, such as the determination of acidic substances, alkaline substances, alcohols, hydrocarbons, aromatic compounds, esters, ketones, and other chemical substances; have been proven to be used in power plant smoke Gas desulfurization lime slurry pH measurement-working principle GLI differential sensor technology uses three measuring electrodes to replace the dual electrodes in traditional pH sensors. The difference between the pH value measured by the measuring electrode and the standard electrode relative to the background electrode of the third solution. The technology has been proven to have precise accuracy, reducing the pollution of the reference electrode connection, and effectively eliminating the closed circuit. cut back
封装Ph_ORP探头_12
封装Ph_ORP探头1
封装Ph_ORP探头31
产品描述
Technical indicators
- typical application
 
Suitable for application in harsh environmental conditions, such as determination of acidic substances, alkaline substances, alcohols, hydrocarbons, aromatic compounds, esters, ketones, and other chemical substances; has been proven to be used in power plant flue gas desulfurization Slurry pH measurement
 
 
 
- working principle
 
GLI differential sensor technology uses three measuring electrodes to replace the two electrodes in traditional pH sensors. The difference between the pH value measured by the measuring electrode and the standard electrode relative to the background electrode of the third solution. The technology has been proven to have precise accuracy, reducing the pollution of the reference electrode connection, and effectively eliminating the closed circuit. Reduced downtime and maintenance time.
 
 
 
-Instrument characteristics
 
● Using differential measurement technology, with excellent accuracy and reliability-replace the two electrodes of the traditional pH sensor with three electrodes
 
● Higher reliability reduces maintenance time and downtime
 
● Built-in preamplifier
 
● Various installation methods
 
● Transmission distance of 914 meters
 
● LCP sensor reduces thermal distortion
 
● Ryton sensor has excellent compatibility with strong alkaline substances
 
● Antimony electrode can withstand hydrofluoric acid
 
● Replaceable salt bridge extends sensor life
 
扫二维码用手机看
未找到相应参数组,请于后台属性模板中添加
Technical index
Range: pH:0-14pH
ORP: -2000~+2000mV
temperature range: -5~95℃
greatest pressure: 6.9Bar
Sensitivity: pH:<0.005pH
ORP: <0.5mV
Wetting material: LCP body, PVDF connection, Viton O-ring, glass electrode
weight: 0.6kg

 