Technical indicators
- typical application
Suitable for application in harsh environmental conditions, such as determination of acidic substances, alkaline substances, alcohols, hydrocarbons, aromatic compounds, esters, ketones, and other chemical substances; has been proven to be used in power plant flue gas desulfurization Slurry pH measurement
- working principle
GLI differential sensor technology uses three measuring electrodes to replace the two electrodes in traditional pH sensors. The difference between the pH value measured by the measuring electrode and the standard electrode relative to the background electrode of the third solution. The technology has been proven to have precise accuracy, reducing the pollution of the reference electrode connection, and effectively eliminating the closed circuit. Reduced downtime and maintenance time.
-Instrument characteristics
● Using differential measurement technology, with excellent accuracy and reliability-replace the two electrodes of the traditional pH sensor with three electrodes
● Higher reliability reduces maintenance time and downtime
● Built-in preamplifier
● Various installation methods
● Transmission distance of 914 meters
● LCP sensor reduces thermal distortion
● Ryton sensor has excellent compatibility with strong alkaline substances
● Antimony electrode can withstand hydrofluoric acid
● Replaceable salt bridge extends sensor life