NPW-160 Total Phosphorus / Total Nitrogen / COD Analyzer

Brand instruments

Brand Instrument

image/y2011712161156

NPW-160 Total Phosphorus / Total Nitrogen / COD Analyzer

image/y2011712161156
Product description
Technical indicators
- working principle:
 
Total phosphorus (according to international GB11893-89): Add potassium persulfate as oxidant to the sample, digest the water sample to be tested at 120 ℃ for 30min, convert the phosphide into phosphate ion, and use the molybdenum blue method to absorb water to detect the water quality The total phosphorus content (measurement wavelength: 700nm)
 
Total nitrogen and potassium persulfate are used as oxidants (according to international GB11893-89): potassium persulfate is used as oxidizer and digested at 120 ° C for 30 minutes to convert the nitrogen-containing compounds into nitrate ions and adjust the pH of the sample solution to 2- 3; The content of total nitrogen in water quality is detected by ultraviolet absorption spectrophotometry (measurement wavelength: 220nm, 275nm turbidity correction: A = A220-A275 × 2)
 
COD (UV): dual-wavelength absorbance photometry (UV 254nm, visible light 546nm)
 
 
 
-Application industry:
 
The NPW-160 total phosphorus / total nitrogen analyzer is mainly used for automatic monitoring of total phosphorus, total nitrogen and COD in surface water, tap water, municipal sewage or other special occasions.
 
 
 
-Instrument features:
 
● Independently designed heating decomposition device
 
● The system can realize wireless transmission
 
● Built-in CF card, data can be stored for a long time
 
● Low operating cost and less secondary pollution
 
● Integrated design simplifies pipeline connection
 
● Advanced multi-wavelength detector can measure three indicators: total phosphorus, total nitrogen, COD (UV)
 
 
 
- Product advantages:
 
● Built-in multi-wavelength detector, one measurement can get two parameters at the same time, including total nitrogen and total phosphorus;
 
● According to JIS standard (heating at 120 oC for 30 min), the same measurement method is adopted; it has good accuracy and stability;
 
● Low operating cost, reduce reagent consumption and waste liquid discharge;
 
● Built-in multiple functions, data storage card (can save one year of measurement data), reagent quantitative sampling system;
 
● Can be equipped with wireless transmission function.
发送反馈
历史记录
Scan the QR code to read on your phone
We could not find any corresponding parameters, please add them to the properties table
Technical index
Measuring range: Total nitrogen, 0 ~ 2mg / L to 200mg / L
Total phosphorus, 0 ~ 0.5mg / L to 20mg / L
Measuring range: ≤± 3% FS
Analysis interval: 1 hour to 6 hours, can be set arbitrarily
Sample conditions: Temperature, 2 ~ 40 ℃
Pressure, 0.02 ~ 0.05MPa
Flow rate, 1 ~ 3L / min
Each analysis sample volume is about 67.5mL
Reagent replenishment interval: 1 month
working environment: Temperature, 2 ~ 40 ℃, 85% relative humidity indoor installation
Analog output: 4 ~ 20 mA, load 600 ohm
Relay: Can set upper and lower limit alarms, fault information, etc.
power supply: 220VAC,50/60HZ
Power consumption: 500VA (maximum) 200VA (average)
Data communication: RS485 串行端口RS485 serial port
Dimensions: 450 × 380 × 1430 mm
weight: About 80Kg

 

Customer message
 Fuguang

Mainly engaged in the distribution of environmental monitoring instruments and meters, as well as the integration and operation and maintenance of environmental monitoring systems; Research and development, production, and sales of environmental monitoring instruments;

Copyright ©2023 Fuzhou Fuguang Water Science&Technology    闽ICP备12013341号-2  Powered by saa 

 Fuguang

Follow Fuguang Water

 Fuguang

Official Taobao Store

Fuzhou Fuguang Water Science&Technology Co.,Ltd

闽ICP备12013341号-2  Powered by saa